Troubleshooting
Potential failures and troubleshooting
5
Maquet PowerLED II
IFU 01811 EN 10
97 / 116
Indicator Description Meaning
Indicator off No fault
Warning indicator Faulty system
Tab.27: Warning indicators
Indicator Description Meaning
Indicator off Maintenance up to date
Maintenance indicator Annual maintenance needed
Tab.28: Maintenance indicators
5.2 Potential failures and troubleshooting
Mechanical
Anomaly Likely cause Corrective action
The sterilisable handle does
not click into place correctly
The locking mechanism is
damaged
Replace the handle
Drift of the system Worn brake(s) Have the brakes replaced by a
trained technician
Incorrect adjustment of the
brake(s)
Have the brakes adjusted by a
trained technician
Device too stiff to manoeuvre Mechanical lock Contact the Getinge technical
department
Tab.29: Mechanical anomalies and malfunctions
Electronics/Optics
Anomaly Likely cause Corrective action
The lighthead does not turn
on.
Power cut Contact your facility’s technical
services
Other reason Contact the Getinge technical
department
The lighthead does not turn
off.
Communication problem Contact the Getinge technical
department
A group of LEDs or one LED
does not come on
The LED board is defective Contact the Getinge technical
department
The light flickers The LED board is defective Contact the Getinge technical
department
Tab.30: Optical anomalies and malfunctions