Use
Daily inspections before use
4
Maquet PowerLED II
IFU 01811 EN 10
43 / 116
4 Use
4.1 Daily inspections before use
NOTE
To ensure that the product used is compliant, various daily visual and functional in-
spections must be performed by trained personnel. It is recommended that records
be kept of the results of these inspections, along with the date and signature of the
person performing them.
Fig.35: Integrity of the device
Integrity of the device
1. Check that the device has not suffered any
impact damage.
2. Check for any chipped or missing paint.
3. If a problem is noted, contact technical
support.
Fig.36: Suspension covers
Suspension covers
1. Check that the spring arm covers are in
the proper position and in good condition.
2. Check that the suspension covers, includ-
ing the one beneath the central shaft, are
in the proper position and in good condi-
tion.
3. If a problem is noted, contact technical
support.
DF
SF
Fig.37: Half-rings
Half-rings on spring arms
1. Check that the half-rings on the spring
arms are in place in their slots.
2. If a problem is noted, contact technical
support.

Table of Contents

Other manuals for Getinge Maquet PowerLED II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet PowerLED II and is the answer not in the manual?

Getinge Maquet PowerLED II Specifications

General IconGeneral
BrandGetinge
ModelMaquet PowerLED II
CategoryMedical Equipment
LanguageEnglish

Related product manuals