4
Use
Daily inspections
38 / 106
VOLISTA
IFU 01781 EN 19
Fig.31: Stability and drift of the system
Stability and drift of the system
1. Operate the device, making several move-
ments in order to swivel the suspension
arms, the spring arms and the lightheads.
Ø The entire system should move easily
and smoothly.
2. Place the system in various positions.
Ø The entire system should remain in the
selected position, without any drift.
3. If a problem is noted, contact technical
support.
Fig.32: Spring arm positioning
Spring arm positioning
1. Place the spring arm in its lowest position,
horizontally and finally in its highest posi-
tion.
2. Check that the spring arm remains in each
of these positions.
3. If a problem is noted, contact technical
support.
Fig.33: Sterilisable handle mount
Sterilisable handle mount
1. Remove the handle mount in place.
Ø Check that it can be removed easily.
2. Reinstall the handle mount on the light-
head.
Ø Check that it can be installed smoothly
and that the handle mount is properly
installed.