Model 574 Programmable Counting System (PCS)
238 - GSE Scale Systems
0.950585
61217%s-524%e P61217. AIN1 NROff
-524
61218%s-2131%e P61218. AIN2 NROff
-2131
61219%s1309%e P61219. AIN4 NROff
1309
61220%s-213%e P61220. AIN8 NROff
-213
61221%s-4697%e P61221. VREF NROff
-4697
61222%s8%e P61222. SN:
8
%z Exit Setup Mode
NOTE: Make sure all installed scales are enabled at
parameters P101 thru P103. Make sure that the
respective scale information is entered into the proper
parameters before calibration.
P110 thru P119 Scale 1
P120 thru P129 Scale 2
P130 thru P139 Scale 3
22.7 SETUP
The calibration parameters for each scale are stored in
EEPROM. Each additional scale requires 31 bytes of
space in the EEPROM. Key in <60001> <SELECT> to
see if you have enough space in EEPROM. If not you
may need to reduce the size of your custom transmits or
macros, or you may wish to expand your EEPROM.
22.8 SCALE ENABLE
Four setup parameters have been added: 101, 102 and
103. Each one of these corresponds to a scale input (1
through 3) and may be set to: 0 - Disabled, 1 - Saved,
and 2 - Enabled. When a scale is disabled, its setup and
calibration is lost. When it's saved, its setup and
calibration information is retained in EEPROM, however
the scale may not be selected from the weigh mode.
When it's enabled, its setup and calibration information
is retained in EEPROM, and it may be selected from the
weigh mode.
22.9 SCALE SETUP
When a scale is saved or enabled, the setup parameters
may be viewed and / or changed through the setup mode.
The following ranges of parameters correspond to the
specified scale inputs:
110 through 119 Scale 1
120 through 129 Scale 2
130 through 139 Scale 3
The parameters are the same for all 3 scales:
1X0 Full Scale
1X1 Display Division
1X2 Zero Track Aperture
1X3 Zero Track Delay
1X4 Motion Band Width
1X5 Motion Delay
1X6 Filter Setting
1X7 Display update rate
1X8 Zero Range
1X9 Linearization Enable
22.10 SCALE CALIBRATION
Scale calibration is performed the same as a standard
Indicator calibration procedure, only when more than
one scale is enabled, you will be prompted "Keyin
Chan#". You then enter the number of the scale you
wish to calibrate. When you finish calibrating it, you
will again be prompted "Keyin Chan#", to allow you to
calibrate the other scales. If you don't want to calibrate
any more scales, press <CLR> to exit. The "QUICK
CAL" works the same way.
22.11 OPERATION
While in the weigh mode, only one scale may be selected
at a time. The GROSS weight, NET weight, TARE,
ZERO, etc., are for the selected scale. Setpoints and
analog outputs, if based on one of these parameters, are
based on the currently selected scale.
It should be noted that the filtering for each scale is
performed by the circuitry in the Multi-Scale option
itself, so a delay is not necessary when switching to
another scale. Zero tracking, however, is active only on
the selected scale.
22.12 SELECTING THE CURRENT
SCALE