GSE Scale Systems - 269
Chapter 29 Troubleshooting Technical Reference Manual
weight. If all appears
correct, refer to the use of the
information parameters,
P61100, and determine the
output (in mv / volt) of the
connected load cell.
32 ADD MORE! The applied weight during
calibration was less than
0.1% of capacity. More
weight than this is required.
Refer to P61100 if this is
incorrect.
33 ReCALReq'd The just completed
calibration is insufficient to
guarantee accurate results
due to either the cal weight
being less than 5% of
capacity or this was the first
calibration of this platform to
this unit and therefore the
coarse gain was adjusted by
the unit.
34 RES> 25K! The current combination of
capacity P110 and increment
P111 result in a resolution
greater than 25,000
graduations. This is simply a
warning in case this was not
intended.
35 RES>100K! The current combination of
capacity P110 and increment
P111 result in a resolution
greater than 100,000
graduations. This is not
allowed and as soon as any
key is pressed the instrument
will jump back into the setup
mode to parameter P110 to
verify the settings.
36 RES< 100! The current combination of
capacity P110 and increment
P111 result in a resolution
less than 100 graduations.
This is simply a warning in
case this was not intended.
37 RES< 1 !! The current combination of
capacity P110 and increment
P111 result in a resolution
less than 1 graduation (i.e.
the increment is greater than
capacity). This is not
allowed and as soon as any
key is pressed the instrument
will jump back into the Setup
Mode to parameter P110 to
verify the settings.
29.6 General Error Messages
41 IDnotUsed! If an ID that has not been
setup (i.e. had its size set to a
non-zero value in P700,
P702, P704, etc.) is accessed
(by pressing 3 <ID> for
example), or if no ID's have
been setup, then this message
will occur.
42 CheckSetup This error occurs if ID usage
P720 is set for Truck In/Out
Weighing and either the ID
type P721 is set for ID #6 and
its size P710 is set for zero,
or insufficient EEPROM is
installed and there isn't room
for any rows of data to be
stored.
99 Can't Set! An attempt to enter a value
for a parameter which is not
field changeable, such as the
serial numbers or the audit
trail counter results in this
message.
- - Cksumerror Upon each power-up, the
unit tests the integrity of
its EPROM. If the result is
not correct this message is
displayed and the unit is
not usable. Verify that the
EPROM (U13 is installed
properly (no bent over pins).
Reseating the EPROM might
take care of the check sum
error.
29.7 Miscellaneous Messages
- - EntryError This error message is the
most commonly used. The
primary causes are entering a
SECTION - 29.6