EasyManuals Logo

GW Instek PSU Series Programming Manual

GW Instek PSU Series
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
PSU Series Programming Manual
100
MIN
Minimum falling current slew rate is
0.001A/msec.
MAX
Maximum falling current slew rate is
rated current divided by 100msec.
Return Parameter
<NR2>
Returns the step current
Example
SOUR:CURR:SLEW:FALL MAX
Sets the falling current slew rate to the maximum.
[:SOURce]:MODE?
Set
Query
Description
Returns the status of the output mode (CC, CV,
Off) of the power supply.
The interface will return CV if the supply is in
Constant Voltage Mode, CC if the supply is in
Constant Current Mode or OFF if the supply
output is off.
Query Syntax
[:SOURce]:MODE?
Return parameter
<string>
Returns the output state as a string, “CC”,
“CV”, “OFF”
Example
:SOUR:MODE?
>CC
The power supply is currently in CC mode.
[:SOURce]:RESistance[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the internal resistance in ohms.
Syntax
[:SOURce]:RESistance[:LEVel][:IMMediate][:AMPLitud
e] {<NR2>(OHM)|MINimum|MAXimum}
Query Syntax
[:SOURce]:RESistance[:LEVel][:IMMediate][:AMPLitud
e]?
Parameter
<NR2>
Resistance in ohms:
0 ohm ~ Rrated = Vrated/Irated

Table of Contents

Other manuals for GW Instek PSU Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU Series and is the answer not in the manual?

GW Instek PSU Series Specifications

General IconGeneral
BrandGW Instek
ModelPSU Series
CategoryPower Supply
LanguageEnglish

Related product manuals