REMOTE CONTROL
111
:SYSTem:CONFigure:OUTPut:PON[:STATe]
Sets the output state at power-on. This is the
equivalent to the F-92 (Output Status when Power
ON) power on configuration settings.
The setting will only be valid after the power has been
cycled.
:SYSTem:CONFigure:OUTPut:PON[:STATe]
{<NR1>|{SAFE|OFF}|{FORCe|ON}|AUTO}
:SYSTem:CONFigure:OUTPut:PON[:STATe]?
The PSU turns on in the same
state the unit was in prior to the
previous shut down. The output is
set to off (default).
The PSU turns on in the same
state the unit was in prior to the
previous shut down. The output is
set to on.
The PSU turns on in the same
state the unit was in prior to the
previous shut down, but with the
same output on/off setting.
The power on output setting is
“SAFE” or “OFF”.
The power on output setting is
“FORCe” or “ON”.
The power on output setting is
“AUTO”.
SYST:CONF:OUTP:PON SAFE
SYST:REB
Set the power-on output state to safe and reboot
the unit to active the setting.