REMOTE CONTROL
125
Example
SYST:COMM:SER:TRAN:PARity?
>1
Indicates that odd parity is used for the UART
connection.
SYST:COMM:SER:TRAN: PARity 1
SYST:REB
Sets the UART parity to odd and reboot the unit to
active the setting.
:SYSTem:COMMunicate:SERial[:RECeive]
:TRANsmit:SBITs
Set
Query
Description
Sets or queries the number of stop bits used for
the UART connection.
The setting will only be valid after the power has been
cycled.
Syntax
Query Syntax
:SYSTem:COMMunicate:SERial[:RECeive]:TRANsmit
:SBITs<NR1>
:SYSTem:COMMunicate:SERial[:RECeive]:TRANsmit
:SBITs?
Parameter/Return
parameter
0
1 stop bit
1
2 stop bits
Example
SYST:COMM:SER:TRAN:SBITs?
>1
Indicates that one stop bit is used for the UART
connection.
SYST:COMM:SER:TRAN: SBITs 1
SYST:REB
Sets the UART stop bits to 2 stop bits and reboot
the unit to active the setting.