REMOTE CONTROL
85
Output Commands
:OUTPut:DELay:ON ........................................................... 85
:OUTPut:DELay:OFF ......................................................... 85
:OUTPut:MODE .................................................................. 86
:OUTPut[:STATe][:IMMediate] ......................................... 86
:OUTPut[:STATe]:TRIGgered ........................................... 86
:OUTPut:PROTection:CLEar ............................................ 87
:OUTPut:PROTection:TRIPped ....................................... 87
:OUTPut:DELay:ON
Set
Query
Description
Sets the Delay Time in seconds for turning the
output on. The delay is set to 0.00 by default.
Syntax
Query Syntax
:OUTPut:DELay:ON {<NR2>|MINimum|MAXimum}
:OUTPut:DELay:ON?
Parameter
<NR2>
0.00~99.99 seconds, where 0=no delay.
Return parameter
0.00
Returns the delay on time in seconds
until the output is turned on.
:OUTPut:DELay:OFF
Set
Query
Description
Sets the Delay Time in seconds for turning the
output off. The delay is set to 0.00 by default.
Syntax
Return Syntax
:OUTPut:DELay:OFF {<NR2> |MINimum|MAXimum}
:OUTPut:DELay:OFF?
Parameter
<NR2>
0.00~99.99 seconds, where 0=no delay.
Return parameter
0.00
Returns the delay off time in seconds
until the output is turned off.

Table of Contents

Other manuals for GW Instek PSU Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU Series and is the answer not in the manual?

GW Instek PSU Series Specifications

General IconGeneral
BrandGW Instek
ModelPSU Series
CategoryPower Supply
LanguageEnglish

Related product manuals