PSU Series Programming Manual
98
Parameter
<NR2>
Current protection level.
Minimum: Depend on the unit type:
if Irated * 0.1 > 5A, then minimum = 5A,
else minimum = Irated * 0.1
Maximum: Irated * 1.1
MIN
Minimum current level.
MAX
Maximum current level.
Return parameter
<NR2>
Returns the current protection level.
Example
SOUR:CURR:PROT:LEV?
+5.000
Returns the minimum possible current level in
amps.
[:SOURce]:CURRent:PROTection:STATe
Set
Query
Description
Turns OCP (over-current protection) on or off.
Syntax
[:SOURce]:CURRent:PROTection:STATe
{<bool>|OFF|ON}
Query Syntax
[:SOURce]:CURRent:PROTection:STATe?
Parameter
OFF | 0
Turns the OCP off.
ON | 1
Turns the OCP on.
Return parameter
<bool>
Returns the over current protection state
in <bool> format.
Example
SOUR:CURR:PROT:STAT OFF
Turns OCP off.
[:SOURce]:CURRent:PROTection:TRIPped
Set
Query
Description
Returns the state of the current protection circuits.
Query Syntax
[:SOURce]:CURRent:PROTection:TRIPped?
Return parameter
<bool>
Returns protection status.