PSU Series Programming Manual
104
SOUR:VOLT:PROT:TRIP?
>0
Indicates that the OVP protection has not been
tripped.
[:SOURce]:VOLTage:SLEWrate:RISing
Sets or queries the rising voltage slew rate. This is
only applicable for CV slew rate priority mode.
[:SOURce]:VOLTage:SLEWrate:RISing
{<NR2>(V)|MINimum|MAXimum}
[:SOURce]:VOLTage:SLEWrate:RISing?
Per step is between 0.001V/msec and
rated voltage divided by 100msec.
Minimum rising voltage slew rate is
0.001V/msec.
Maximum rising voltage slew rate is
rated voltage divided by 100msec.
Returns the slew rate in V/msec.
SOUR:VOLT:SLEW:RIS MAX
Sets the rising voltage slew rate to its maximum.
[:SOURce]:VOLTage:SLEWrate:FALLing
Sets or queries the falling voltage slew rate. This is
only applicable for CV slew rate priority mode.
[:SOURce]:VOLTage:SLEWrate:FALLing
{<NR2>(V)|MINimum|MAXimum}
[:SOURce]:VOLTage:SLEWrate:FALLing?