REMOTE CONTROL
135
Queries the Standard Event Status (Event)
register. The Event Status register is cleared after
it is read.
Returns the bit sum of the Standard
Event Status (Event) register and clears
the register.
Queries the manufacturer, model name, serial
number, and firmware version of the PSU.
Returns the instrument identification as
a string in the following format:
GW-INSTEK, PSU-20-76, TW123456,
01.00.20110101
Manufacturer: GW-INSTEK
Model number: PSU-20-76
Serial number: TW123456
Firmware version: 01.00.20110101
The *OPC command sets the OPC bit (bit0) of the
Standard Event Status Register when all current
commands have been processed.
The *OPC? Query returns 1 when all the
outstanding commands have completed.