Source Commands
[:SOURce]:CURRent[:LEVel][:IMMediate]
[:AMPLitude] ......................................................................... 95
[:SOURce]:CURRent[:LEVel]:TRIGgered
[:AMPLitude] ......................................................................... 96
[:SOURce]:CURRent:LIMit:AUTO .................................. 96
[:SOURce]:CURRent:PROTection:DELay ...................... 97
[:SOURce]:CURRent:PROTection[:LEVel] ..................... 97
[:SOURce]:CURRent:PROTection:STATe ...................... 98
[:SOURce]:CURRent:PROTection:TRIPped .................. 98
[:SOURce]:CURRent:SLEWrate:RISing ........................... 99
[:SOURce]:CURRent:SLEWrate:FALLing ....................... 99
[:SOURce]:MODE? ............................................................ 100
[:SOURce]:RESistance[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 100
[:SOURce]:VOLTage[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 101
[:SOURce]:VOLTage[:LEVel]:TRIGgered
[:AMPLitude] ....................................................................... 101
[:SOURce]:VOLTage:LIMit:AUTO ................................ 102
[:SOURce]:VOLTage:LIMit:LOW .................................. 102
[:SOURce]:VOLTage:PROTection[:LEVel] .................. 103
[:SOURce]:VOLTage:PROTection:TRIPped ................ 103
[:SOURce]:VOLTage:SLEWrate:RISing ........................ 104
[:SOURce]:VOLTage:SLEWrate:FALLing .................... 104
[:SOURce]:CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps. For
externally set current levels (from the analog
control connector) the set current level is returned.
Syntax
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NR2>(A)|MINimum|MAXimum}
Query Syntax
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]
?
Parameter/Return
<NR2>
0~105% of the rated current output level.

Table of Contents

Other manuals for GW Instek PSU Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU Series and is the answer not in the manual?

GW Instek PSU Series Specifications

General IconGeneral
BrandGW Instek
ModelPSU Series
CategoryPower Supply
LanguageEnglish

Related product manuals