REMOTE CONTROL
121
The setting will only be valid after the power has been
cycled.
Syntax
Query Syntax
:SYSTem:COMMunicate:LAN:DHCP
{<bool>|OFF|ON}
:SYSTem:COMMunicate:LAN:DHCP?
Parameter
OFF | 0
DHCP off
ON | 1
DHCP on
Return parameter
<bool>
Returns the DHCP status.
Example
SYST:COMM:LAN:DHCP ON
SYST:REB
Sets the DHCP active and reboot the unit to active
the setting.
:SYSTem:COMMunicate:LAN:DNS
Set
Query
Description
Sets or queries the DNS address.
The setting will only be valid after the power has been
cycled.
Syntax
Query Syntax
:SYSTem:COMMunicate:LAN:DNS <string>
:SYSTem:COMMunicate:LAN:DNS?
Parameter/Return
<string>
DNS in string format ( “mask”)
Applicable ASCII characters: 20H to 7EH
Example
SYST:COMM:LAN:DNS “172.16.1.252
SYST:REB
Sets the DNS to 172.16.1.252 reboot the unit to
active the setting.
:SYSTem:COMMunicate:RLSTate
Set
Query
Description
Enables or disables local/remote state of the
instrument.

Table of Contents

Other manuals for GW Instek PSU Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU Series and is the answer not in the manual?

GW Instek PSU Series Specifications

General IconGeneral
BrandGW Instek
ModelPSU Series
CategoryPower Supply
LanguageEnglish

Related product manuals