PSU Series Programming Manual
112
:SYSTem:CONFigure:PROTection:RECovery
Sets or queries how the OHP, FAN, AC-FAIL, and
SD alarms are cleared.
:SYSTem:CONFigure:PROTection:RECovery
{SAFE|AUTO}
:SYSTem:CONFigure:PROTection:RECovery?
The output is not turned on
automatically when the cause of the
alarm is fixed.
The output is turned on automatically
when the cause of the alarm is fixed.
Sets or queries the unit operation mode.
This is the equivalent to the F-93 (Master/Slave
Configuration) power on configuration settings.
:SYSTem:CONFigure:MSLave { <NR1> }
:SYSTem:CONFigure:MSLave?
Parameter/
Return parameter
Master/with 1 unit in Parallel (total:2
units)
Master/with 2 units in Parallel (total: 3
units)
Master/with 3 units in Parallel (total: 4
units)
SYST:CONF:MSL 2
SYST:REB
Set to Master (with 2 slave units in Parallel) and
reboot the unit to active the setting.