132
Global Commands
:GLOBal:CURRent[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 132
:GLOBal:VOLTage[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 132
:GLOBal:OUTPut[:STATe][:IMMediate] ....................... 133
:GLOBal:PRESet ................................................................ 133
:GLOBal:CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Description
Sets current value for all units when using the
(Legacy) Multi-Drop mode.
Syntax
GLOBal:CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NR2>(A)|MINimum|MAXimum}
Parameter
<NR2>
0~105% of the rated current output level.
MIN
Minimum current level.
MAX
Maximum current level.
Example
GLOB:CURR MAX
Sets the current to the maximum for all units.
:GLOBal:VOLTage[:LEVel][:IMMediate]
[:AMPLitude]
Set
Description
Sets voltage value for all units when using the
(Legacy) Multi-Drop mode.
Syntax
:GLOBal:VOLTage[:LEVel][:IMMediate][:AMPLitude]
{<NR2>(V)|MINimum|MAXimum}
Parameter
<NR2>
0~105% of the rated voltage output level
MIN
Minimum voltage level.
MAX
Maximum voltage level.

Table of Contents

Other manuals for GW Instek PSU Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU Series and is the answer not in the manual?

GW Instek PSU Series Specifications

General IconGeneral
BrandGW Instek
ModelPSU Series
CategoryPower Supply
LanguageEnglish

Related product manuals