REMOTE CONTROL
103
Parameter/Return
<NR2>
0 ~ the present setting voltage
MIN
Minimum allowed voltage level
MAX
Maximum allowed voltage level
Example
SOUR:VOLT:LIM:LOW MAX
Sets the UV> level to its maximum.
[:SOURce]:VOLTage:PROTection[:LEVel]
Set
Query
Description
Sets or queries the overvoltage protection level.
Syntax
[:SOURce]:VOLTage:PROTection[:LEVel]
{<NR2>(V)|MINimum|MAXimum}
Query Syntax
[:SOURce]:VOLTage:PROTection[:LEVel]?
Parameter/Return
<NR2>
Minimum: Depends on the unit type:
if Vrated * 0.1 > 5V, then Minimum = 5V,
else Minimum = Vrated * 0.1
Maximum: Vrated * 1.1
MIN
Minimum OVP level
MAX
Maximum OVP level
Example
SOUR:VOLT:PROT:LEV MAX
Sets the OVP level to its maximum.
[:SOURce]:VOLTage:PROTection:TRIPped
Set
Query
Description
Sets or queries the overvoltage protection level.
Query Syntax
[:SOURce]:VOLTage:PROTection:TRIPped?