REMOTE CONTROL
99
SOUR:CURR:PROT:TRIP?
>0
The protection circuit has not been tripped.
[:SOURce]:CURRent:SLEWrate:RISing
Sets or queries the rising current slew rate. This is
only applicable for CC slew rate priority mode.
[:SOURce]:CURRent:SLEWrate:RISing
{<NR2>(A)|MINimum|MAXimum}
[:SOURce]:CURRent:SLEWrate:RISing?
Per step is between 0.001A/msec and
rated current divided by 100 msec.
Minimum rising current slew rate is
0.001A/msec.
Maximum rising current slew rate is
rated current divided by 100msec.
Returns the step current in amps.
SOUR:CURR:SLEW:RIS?
0.950
Sets the rising current slew rate to 0.950 A/ms.
[:SOURce]:CURRent:SLEWrate:FALLing
Sets or queries the falling current slew rate. This is
only applicable for CC slew rate priority mode.
[:SOURce]:CURRent:SLEWrate:FALLing
{<NR2>(A)|MINimum|MAXimum}
[:SOURce]:CURRent:SLEWrate:FALLing?
Per step is between 0.001A/msec and
rated current divided by 100 msec.