Page #99 background image
REMOTE CONTROL
99
Example
SOUR:CURR:PROT:TRIP?
>0
The protection circuit has not been tripped.
[:SOURce]:CURRent:SLEWrate:RISing
Set
Query
Description
Sets or queries the rising current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax
Query Syntax
[:SOURce]:CURRent:SLEWrate:RISing
{<NR2>(A)|MINimum|MAXimum}
[:SOURce]:CURRent:SLEWrate:RISing?
Parameter
<NR2>
Per step is between 0.001A/msec and
rated current divided by 100 msec.
MIN
Minimum rising current slew rate is
0.001A/msec.
MAX
Maximum rising current slew rate is
rated current divided by 100msec.
Return parameter
<NR2>
Returns the step current in amps.
Example
SOUR:CURR:SLEW:RIS?
0.950
Sets the rising current slew rate to 0.950 A/ms.
[:SOURce]:CURRent:SLEWrate:FALLing
Set
Query
Description
Sets or queries the falling current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax
[:SOURce]:CURRent:SLEWrate:FALLing
{<NR2>(A)|MINimum|MAXimum}
Query Syntax
[:SOURce]:CURRent:SLEWrate:FALLing?
Parameter
<NR2>
Per step is between 0.001A/msec and
rated current divided by 100 msec.

Table of Contents

Other manuals for GW Instek PSU Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSU Series and is the answer not in the manual?

GW Instek PSU Series Specifications

General IconGeneral
BrandGW Instek
ModelPSU Series
CategoryPower Supply
LanguageEnglish

Related product manuals