3.1.6 Health Notes
This system produces hazardous voltages that can cause shocks, burns or death.
During operation high electric and magnetic fields can occur. The fields can affect to your health.
Different measures can help here to reduce the risk (example: shielding against the fields,
increase of the distance to the high voltage system, duration of the test, lower voltage/ current
levels).
Some system configurations produce loud or monotonous noises. Different measures can help
here to reduce the noise (example: acoustic shielding, ear protection, increase of the distance to
the high voltage system, duration of the test, lower voltage/ current levels).
People with heart pacemakers should not be in the vicinity of the system during operation.
3.1.7 Device Particular safety notes
The earth connection of the AXOS must be connected to a good earth.
If trigger is pressed without connection to earth, electrical shock may occur!
Remove all external connection cables before removing any covers.
Do not open any system modules, they contains no user replaceable parts.
Never touch the Equipment Under Test (EUT), when the AXOS is operating.
Establish a safety barrier around the test setup (Close the circuit pin 4 and pin 5 in
Table 5-3).
If any part of the AXOS is damaged or it is possible that damage has occurred, for
example during transportation, do not switch on the unit.
Before changing the mains fuse, remove the mains power cord. Fuses should only
be replaced with the same type and value.
When applying Magnetic field, do not touch the coil when test is running. Only
direct output HI and COM must be used to connect the coil. The coil loop (from HI
till COM) must be closed..
Never connect a supply voltage directly to the HI/COM surge outputs. This will
damage the unit!!!. Always use a suitable CDN