6. Troubleshooting
If the problem persists after trying the corrective action listed for the relevant advisory
number in Table 6.2, please contact technical support.
Table 6.2: List of Advisories.
No. Cause Corrective Action
2
Blood sample not detected due to
inadequate amount of blood used
Repeat the test with a new cartridge. Be
sure to follow the on-screen instructions
6
No clot detected. Can be due to
a very sick animal or the tube not
filled correctly
Repeat the test with a new cartridge. Be
sure to follow the on-screen instructions
7 Flow too fast in the cartridge Repeat the test with a new cartridge. Be
sure to follow the on-screen instructions
8 Flow too slow in the cartridge Repeat the test with a new cartridge. Be
sure to follow the on-screen instructions
9 Sample added too early
Wait for the message Add sample and
wait before adding sample to the cartridge.
Be sure to follow the on-screen instruc-
tions
10
Malfunction of instrument compo-
nents
Turn off and restart the instrument
11
On-going test terminated by the
user
Repeat the test with a new cartridge
12 Ambient temperature too low
Restore recommended ambient condi-
tions for the instrument
13 Ambient temperature too high
Restore recommended ambient condi-
tions for the instrument
14–19 Light detection issue
Repeat the test with a new cartridge and
ensure that no foreign objects are in the
cartridge slot
20 No cartridge detected
Make sure that the cartridge is fully in-
serted in the cartridge slot
22–23 Light detection issue
Repeat the test with a new cartridge and
ensure that no foreign objects are in the
cartridge slot
24 LIS is not responding Check LIS and Ethernet cable
Continued on next page