125
8.6 Message Reference
Resistance Measurement Range
Command
Query
Response
[:SENSe:]RESistance:RANGe <Expected measurement value>
<Expected measurement value> = 0 to 110E+6
[:SENSe:]RESistance:RANGe?
<Measurement Range (NR3)>
<Measurement Range (NR3)> = 20.0000E-3, 200.000E-3, 2000.00E-
3, 20.0000E+0, 200.000E+0, 2000.00E+0, 20.0000E+3, 110.000E+3,
1100.00E+3, 11.0000E+6 or 110.0000E+6
Description
Command Enter the expected measurement value. The instrument is set to the
most suitable range for measuring the given numerical value data.
Query Queries the measurement range setting.
Example
Command
RES:RANG 123
Sets the Resistance function to the 200 range.
Note
The following HIOKI 3227 command can be used, but the format of the response
message is different.
:RESIstance:RANGe
Set and Query the Auto-Ranging Setting
Low-Power Resistance Measurement Range
Syntax
Command
Query
Response
[:SENSe:]LPResistance:RANGe:AUTO <1, 0, ON or OFF>
[:SENSe:]LPResistance:RANGe:AUTO?
<ON or OFF>
Example
Command
LPR:RANG:AUTO ON
Resistance Measurement Range
Syntax
Command
Query
Response
[:SENSe:]RESistance:RANGe:AUTO <1, 0, ON or OFF>
[:SENSe:]RESistance:RANGe:AUTO?
<ON or OFF>
Example
Query
Response
RES:RANG:AUTO?
OFF
Note
The following HIOKI 3227 command can be used, but the format of the response
message is different.
:RESIstance:AUTO