Notes on Use
______________________________________________________________
DANGE
R
To avoid the risk of electric shock and damage to the
product
Do not apply more than 450 V AC or DC between the
analog inputs and product, or between the analog
inputs.
Use only the specified Model 9418-15 AC ADAPTER
(SA145A-1240V-6, SINO AMERICAN). AC adapter input
voltage range is 100 to 240 V AC (with
10% stability)
at 50/60 Hz. To avoid electrical hazards and damage to
the product, do not apply voltage outside of this range.
Terminal inputs should never exceed the limits
indicated below.
Input/output terminals Maximum input voltage
Analog input 450 V AC/DC
EXT TRIG -5 V to +10 V
TRIG OUT
-20Vto+30V
500 mA max, 200 mW max
The maximum voltage applied to the probe and cables
connecting to the analog input of the product must
never exceed the limits in the following table.
Notwithstanding these limits, the voltage at the analog
input must not exceed 450V AC or DC.
Maximum input voltage
9197 CONNECTION CORD
500 V AC/DC
9217 CONNECTION CORD
300 V AC/DC
9322DIFFERENTIAL
PROBE
1000 V AC, 2000 V DC (CAT ll)
600 V AC/ DC (CAT lll)
Notes on Use
Follow these precautions to ensure safe operation and to
obtain the full benefits of the various functions.