8.8 Internal Trigger
______________________________________________________________
Flashing cursor
Flashing cursor
NOTE
The settable standard voltage and control range are limited by
the voltage range setting.
4. Move the flashing cursor to the position shown in the
figure below and set the control range (within
0.5 to 5
divisions of measurement range) using the
buttons.
5. When 50- or 60-Hz Sine Wave was selected in step 3, set
the Standard Voltage.
Move the flashing cursor to the position shown in the
figure below and set the standard voltage using the
buttons.
When measuring 100-V commercial lines:
Set the voltage axis range to 20 V/DIV or 50 V/DIV and
standard voltage to 100 Vr.
When measuring 200-V commercial lines:
Set the voltage axis range to 50 V/DIV or 100 V/DIV and
standard voltage to 200 Vr.