165
_____________________________________________________________________
8.9 Timer Trigger Function
______________________________________________________________
Flashing cursor
4. Move the flashing cursor to
Start
and set the start time
using the
buttons.
To set the current time, press the
button.
Also set the Stop Time and Interval in the same way.
Note on use for the timer trigger
Set the present time on the System Screen first, then set the
timer trigger.
Set the start time and end time to a point after the pressing of
the
START
button.
When the trigger mode is set to
SINGLE
, only one trigger event
is valid for the start time. The settings for time interval and end
time are invalid (when timer trigger only is ON).
To perform recording at regular intervals, establish the following
settings.
Trigger mode: repeat. Other trigger sources: all OFF
Between end of recording and standby, triggers are invalid while
processing is taking place (auto save, auto print, waveform
display processing, calculation) and therefore data may not be
recorded at regular intervals with some measurement setting.
When trigger sources are set to OR:
All trigger sources are valid. If other trigger sources have been
set, triggering can also occur before the start time or after the
end time.
When trigger sources are set to AND:
Measurement is carried out from the start time to the end time.
Triggering occurs once at the preset intervals if the conditions
for the other trigger sources are also met at these points.
If the interval has been set to 0 s, triggering occurs at any point
between the start and end time, if the conditions for the other
trigger sources are met.