10.2 Screen Key Operation Settings
201
Chapter 10 System Environment Settings
10
Select the display method for the horizontal axis at
the bottom of the screen. This setting also deter-
mines the time display for data saved in CSV format.
To help avoid operator errors, confirmation messages can be displayed when mea-
surement is manually started or stopped.
Selecting the Horizontal (Time) Axis Display
1d2h3m4s
(Time)
08-07-10 12:10:30
(Date/Time)
10
(Data Point)
Select from the setting options of
[Time Disp] (Time Value Display).
Setting options (. default setting)
Time
Displays the time span from the start of
measurement. If triggering is enabled, the
displayed time span is from the last trigger
event.
Date Displays the date and time every ten divi-
sions.
Data Pts
(Data points)
Displays the number of data points from the
start of measurement. If triggering is en-
abled, the displayed data points begin from
the last trigger event.
Display of Start/Stop Confirmation Messages
Select from the setting options of
[Start/Stop message].
Setting options (. default setting)
Off Confirmation messages are not displayed.
Pressing START and STOP start and stop mea-
surement immediately.
On
Confirmation messages are displayed. To start
or stop measurement, move the cursor to "Yes"
and press ENTER.