4.6 Setting the Screen Display
89
Chapter 4 Setting the Measurement Conditions (When you want to customize the settings)
4
The screen display of recording data can be changed. The screen display can also
be changed for a recorded waveform before measurement.
The waveform display can be expanded and com-
pressed in the horizontal axis direction. This setting can
also be changed after the waveform has been acquired.
The setting items differ depending on the function.
When High-speed Function (Time Base Setting)
Change the display magnification
4.6 Setting the Screen Display
Expanding and Compressing in the Horizontal Axis (Time Axis)
Direction
Normal
Display
Zoomed
Display
1
2
(Open/confirm the setting information.)
Select from the setting options of [Zoom].
Setting options (∗. default setting)
x10, x5, x2, x1∗, x1/2, x1/5, x1/10, x1/20, x1/50, x1/100,
x1/200, x1/500, x1/1000, x1/2000
The setting can also be changed on the Waveform screen.
Example: When the time base range is 100 µs/div
The value per division is as follows depending on the
magnification.
• when x10: 100μs/div÷10 = 10μs/div
• when x1/100: 100
μs/div x 100 = 10ms/div
Select the [Setting] screen.