Press the START key.
(The green LED lights.)
[Storing...] is displayed on the screen during measurement.
When measuring using the trigger function, the timing for starting
measurement and starting recording (starting acquiring of data to in-
ternal buffer memory) differs.
Start Measurement
IMPORTANT
When the START key is pressed, the start confirmation message appears.
(You can set the message not to be displayed (p. 201).)
If measurement is resumed after it has been stopped, the measurement
data recorded to the internal buffer memory will be overwritten. Save any
important data to a CF card or USB flash drive before you resume mea-
surement.
When automatically saving data
When you want to save the data to storage media simultaneously with measure-
ment, check the following items before you start recording.
• Are the auto save settings configured properly? (p. 102)
• Is the CF card or USB flash drive inserted properly? (p. 46)
• Is there sufficient free space on the CF card or USB flash drive? (p. A15)
When [Repeat] is set to [Single] (default setting), measurement fin-
ishes when the data is read once for the set recording length (or re-
cording time portion).
When [Repeat] is set to [Repeat] or [Cont (Continuous Recording)]
is set to [On] (Real-time Function only), press the STOP key to end
measurement.
If you press the STOP key part way through measurement, the stop
confirmation message appears. (You can set the message not to be
displayed (p. 201).)
Select [Yes] to force measurement to stop.
Finish Measurement