4.4 Using Real-time Function (Fluctuation Recording)
85
Chapter 4 Setting the Measurement Conditions (When you want to customize the settings)
4
Maximum
value
Minimum
value
: Oversampled data
: Saved data (two data items, maximum value and
minimum value)
Recording
interval
Oversampling data (1MS/s)
: Waveform output to the screen or recording paper
Envelope Values
One sampling data for during
[Envelope] recording consists
of two data items, the maximum
value and minimum value, of
the measurement values ob-
tained as a result of performing
oversampling during the set re-
cording interval.
When displaying them on the
screen or printing them with the
printer, they will be displayed or
printed as a shaded band.
When saved to external storage
media, the two data items, the
maximum value and minimum
value, are saved for one mea-
surement time.
If real-time saving, [Envelope] function, and [Numerical Calc.] are set at
the same time for the Real-time Function, the recording interval cannot
be set to 100 µs and 200 µs.