Hioki PW3390 - 6.1 Initializing the Instrument (System Reset)

Hioki PW3390
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.7 Trend Function
122
Scale 1/8, 1/4, 1/2, ×1, ×2, ×5, ×10, ×20, ×50, ×100, ×200, ×500
Setting the scale
Press the key or the
key to select the scale.
Press the key for the
AUTO setting.
Press the key for the
MANUAL setting.
Select [Scale] for the
graph number you
wish to change.
Setting the scale to [×1] causes the value per division on the vertical axis to be 25% of the
full-scale value of each plotted item.
For more information about the relationship between graphed parameters and referenced
full-scale values, see "8.3.3 Output Level" (p.169).
When the scale is set to [AUTO], the scale value will be set such that the maximum and
minimum values for the waveform being graphed fit in the available space.
When the scale is set to [MANUAL], you can set the maximum and minimum values that
will be displayed on the trend graph.
Setting the trend graph reference position
Select [Base Div]
for the graph number
you wish to change.
Base Div: -4 to +4
Use F keys to select
Changing the reference position while using the [AUTO] scale setting may cause a scale
value that differs from the scale value selected for the initial position (reference position 0)
to be selected due to a change in the domain that can be graphed.
•The [Base Div] setting is not available while using the [MANUAL] scale setting.
Only values between the set maximum and minimum values are graphed.