5.7 Trend Function
123
5
Chapter 5 Operating Functions
Setting the value that corresponds to the reference position (percent of full scale)
The formula displayed above allows you to check the value of the plotted item that corre-
sponds to the reference position.
The [Base %] setting is not available while using the [AUTO] scale setting. The value that
corresponds to the reference position is determined automatically based on the maximum
and minimum values for the waveform being graphed.
•The [Base %] setting is not available while using the [MANUAL] scale setting.
Only values between the set maximum and minimum values are graphed.
Select [Base %] for
the graph number you
wish to change.
Use F keys to set.
Base%:
-300.00% to +300.00%
Setting the trend graph maximum and minimum values
The [MAX] and [MIN] settings are not available when using a scale setting other than
[MANUAL].
No trend graph will be generated if the [MAX] value is less than the
[MIN] value, or if the
[MIN] value is greater than the [MAX] value.
Select [MAX] or [MIN]
for the graph number
you wish to change.
Open a value entry dialog box with
the key or key.