4.3 Integration Value Observation
70
Integration is performed for the specified duration, and stops when the timer expires. Calculation results
are held constant when the timer stops.
If auto-saving is enabled, integration values are saved to CF card when integration starts and stops. If an
interval time is also specified, total integration values up to that point are saved at each interval.
See "7.5.2 Auto-Saving Measurement Data" (p. 141)
Timer-Controlled Integration
Setting Procedure
Integration stops when the timer expires (or the clock stop time occurs). When this occurs
before the end of an interval, the last interval is ignored.
Setting range is 10 seconds (“0 hour 0 min 10 sec”) to “9999 hour, 59 min 59 sec”.
If clock timing control is set longer than the timer setting, integration starts at the clock start
time and stops when the timer has expired (the clock stop time is ignored).
Pressing before the timer expires during timer controlled integration stops integration
and retains the integration values. In this instance, pressing again resumes integration
and continues for the set timer duration (additional integration).
Display the
[Time] page
to enable (set to [ON])
To interrupt integration:
Press again.
Use the F keys to set the timer
Press to start integration, which stops
automatically after the set time elapses.
Select
[Timer mode]
Select a
[Timer setting] digit
Lights green