4.3 Integration Value Observation
71
4
Chapter 4 Viewing Measurement Values
After pressing , the instrument waits until the specified clock start time. Integration then begins and
continues until the specified clock stop time.
If auto-saving is enabled, integration values are saved to CF card at the specified start and stop times. If
an interval time is also specified, total integration values up to that point are saved after each interval.
Clock-Controlled Integration
Setting Procedure
• Clock control settings are in 1-minute units.
• Clock setting years are AD (Christian Era), and 24-hour time (e.g., December 6th 2017, 10:16
PM appears as 2017-12-06 22:16)
• If a specified clock time has already passed, clock control is considered to be disabled (OFF).
• When integration is interrupted during clock timing control, clock control is disabled (OFF).
• If clock timing control is set longer than the timer setting, integration starts at the clock start
time and stops when the timer has expired (the clock stop time is ignored).
• Integration stops after 9999 hours, 59 minutes, and 59 seconds if the time between clock start
and stop times is set longer than that.
Display the
[Time] page
Select [Real time]
Select a [Start time]
and [Stop time] digit
(automatically starts and stops
at the specified clock times)
To abort automatic control (while waiting):
Press again.
Use the F keys to set the timer
to enable (set to [ON])
During wait: lights yellow
At the start time: lights green