4.5 Viewing Waveforms
80
Displaying Separate Channel Waveforms
Select [CH]
(Separate Channel)
Voltage Scale
(Positive-only display)
Current Scale
(Negative-only display)
RMS (rms)
Waveform Display En-
able/Disable (p. 81)
Negative Peak
Value (pk-)
Positive Peak
Value (pk+)
Waveform color ap-
pears when enabled
(ON).
• Waveforms and numerical measurement values displayed at the right are not synchronized
with measurement timing.
• Displayed waveform values are not the calculated RMS and peak numerical values.
• The vertical axis of the waveform is displayed as a percentage of the full-scale range of each
channel, so the amplitudes of different channels are not directly comparable.
• To display waveforms starting from zero amplitude, see "Zero-Crossing Filter Setting Proce-
dure" (p. 57).
• Pressing the HOLD key will trigger the HOLD state. However, display data updates will not
function for the waveform display. See "5.3.1 Data Hold Function" (p. 112).