Measure the Clock Period Manually
Time and voltage measurements can be made manually by turning the
T Markers field on.
1
Press the Marker MENU key. Marker is printed in white on the MENU
key.
2 Highlight the T Markers Off field and press the Select key. Highlight On
in the pop-up menu, and press the Select key.
3 Highlight the Trig to X field. Using the knob, move the X Marker to the
lowest point on the negative pulse.
You may need to expand or compress the waveform, using the s/Div field to
see a complete period of the waveform on the display.
4
Highlight the Trig to O field. Using the knob move the O Marker to the
same point on the next negative edge after the X Marker.
5 Read the period of the clock in the Tx to To field.
Measuring the Clock Period Manually with the X and O Markers
Highlight this field
and select On.
Highlight this field and
position the X Marker.
Highlight this field and
position the O Marker.
Read the clock
period in this field.
Using the Oscilloscope
Measure the Clock Period Manually
7-8
Get other manuals https://www.bkmanuals.com

Table of Contents

Other manuals for HP 1660 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 1660 Series and is the answer not in the manual?

HP 1660 Series Specifications

General IconGeneral
CategoryLogic Analyzer
Glitch TriggerYes
Pattern TriggerYes
State AnalysisYes
Transitional TimingYes
Trigger ModesEdge, Pattern, Glitch

Related product manuals