3.
To
create a test sequence, enter the parameters for the measurement that you wish to make.
For this example, a SAW filter measurement is set up with the following parameters:
The above keystrokes will create a displayed list as shown:
Start of Sequence
RECALLPRSTSTATE
~UIS:
PMD
s21
(R/R)
LOG MAG
CENTER
134 M/u
SPAN
50
M/u
SCALE/DIV
AUTO SCALE
4. To complete the sequence creation, press:
Caution
When you create a sequence, the analyzer stores it in volatile memory where
it will be lost if you switch off the instrument power (except for sequence
#6
which is stored in the analyzer non-volatile memory). However, you may store
sequences to a floppy disk.
Running a Sequence
To run a stored test sequence, press:
Preset]
and the softkey labeled with desired sequence number
or, press:
,,
i_
_
..~...
m
~~~~~~~,
md
the
softiey
labeled
ti*
the
desired
sequence
nmber*
Stopping a Sequence
To stop a sequence before it has finished, press
m.