Calibrating for Noninsertable Devices
A test device having the same sex connector on both the input and output cannot be connected
directly into a transmission test configuration. Therefore, the device is considered to be
noninsertable,
and one of the following calibration methods must be performed:
n
adapter removal
n
matched adapters
w
modify the
cal
kit thru definition
NETWORK ANALYZER
REFERENCE
PORT 1
REFERENCE
pg645e
Figure
5-10.
Noninsertable Device
640
Optimizing Measurement Results