11.
lb
perform a one sweep power meter calibration over the IF frequency range at 0
dBm,
press:
Note
Because power meter calibration requires a longer sweep
time,...yyou
may want
.“‘.
.c.
..
to reduce the number of points before pressing
~~~;~~~~.
After the
power meter calibration is finished, return the number of points to its
originaI
value and the analyzer
wilI
automatically interpolate this calibration.
12.
13.
‘lb
calibrate the R channel over the IF range, press:
Once completed, the display should read 0
dBm.
Make the connections as shown in Figure 3-6 for the one-sweep power meter calibration
over the RF
range.
POWER METER
NETWORK ANALYZER
0
.
I
--Ic
[
POW-N&OR
3
dB
pg626e
Figure 3-6.
EXTERNAL
LO
SOURCE
Connections for a One-Sweep Power Meter Calibration for Mixer Measurements
14.
To
set the frequency offset mode Lo frequency from the
analyzer,
press:
Making Mixer Measurements
34