EasyManua.ls Logo

HP 8753E - Connections for a High Dynamic Range Swept if Conversion Loss Measurement

HP 8753E
699 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NETWORK ANALYZER
1
IO
dB
FILTER
EXTERNAL
LO
SOURCE
pg630e
Figure 3-12.
Connections for a High Dynamic Range Swept IF Conversion Loss Measurement
14. To set the frequency offset mode LO frequency, press:
15. To select the converter type and low-side
Lo
measurement configuration, press:
In this low-side
Lo,
down converter measurement, the analyzer’s source frequency range
wiII
be offset higher than the receiver frequency range. The source frequency range can be
determined from the following equation:
receiver frequency range (100
-
1000 MHz) +
Lo
frequency (1500 MHz)
=
1.6-2.5
GHz
16.
To
view the conversion loss in the best vertical resolution, press:
F’igure
3-13 shows the conversion loss of this low-side
Lo,
mixer with output filtering.
Notice that the dynamic range from the pass baud to the noise floor is
welI
above the
dynamic range
limit
of the R Channel. If the mixer under test
also
contained amplification,
then this dynamic range would have been even greater due to the conversion gain of the
mixer.
Making Mixer Measurements
3-15