15. To select the converter type and a high-side 
Lo
 measurement configuration, press:
&&f&g
:
..,.
 
.,.v.
.;.
:.
~6~,~~.~~~~.
/....~~/...~ 
 . . . . 
T.,,
..: ;L?.::<:
 . . . . . . . . 
,,,,......
&&Jal~
.i
Notice in this high-side LO, down conversion configuration, the analyzer’s source is 
actuahy
sweeping backwards, as shown in Figure 3-7. The measurement setup diagram is shown in
Figure 3-8.
100 
MHz
350
 
550
650
900
1
 
GHz
Figure 3-7. 
Diagram
 of Measurement Frequencies
NETWORK ANALYZER
FREO
 
OFFS
ON
 
off
LO
MENU
DOWN
CONVERTER
UP
CONVERTER
RF
 
>
 
LO
RF
 < 
LO
VIEW
MEASURE
RETURN
pg6155d
FIXED  
LO:
1
 
GHz
LO
 
POWER.
13
 
dBm
pg627e
Figure 3-8. Measurement Setup from Display
16. To view the measurement trace, press:
17. To perform a one-sweep power meter calibration over the RF frequency range, press:
Note
Do 
not
 
reduce the number of points to perform this power meter calibration.
Reducing the number of points will turn off the receiver calibration.
The analyzer is now displaying the conversion loss of the mixer calibrated with power
meter accuracy.
3-10
Making Mixer Measurements