Caution
To prevent connector damage, use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.
NETWORK ANALYZER
Figure 3-21.
IO
dB
MIXER
UNDER TEST
3
dB
EXTERNAL
LO SOURCE
pg635e
Connections for the Second Portion of Conversion Compression Measurement
8.
To
set the frequency offset mode
Lo
frequency, press:
9. To select the converter type, press:
10.
‘Ib
select a low-side
Lo
measurement
conflguration,
press:
In this low-side Lo, up converter measurement, the analyzer source frequency is offset lower
than the receiver frequency. The analyzer source frequency can be determined from the
following equation:
receiver frequency (800 MHz)
-
Lo
frequency (600 MHz) = 200 MHz
330 Making Mixer Measurements