n
time domain (Option 010 Only)
Use
‘lhble
6-l
to determine the
minimum cycle time for the listed measurement parameters.
The values listed represent the
minimum time required for a
CW
time measurement with
averaging off.
‘12l.ble
6-1. Minimum Cycle Time (in seconds)
Number of Points
IF Bandwidth
6oooHz
3700 Hz
aoooIIz
11
0.0025 0.0041 0.0055
51
0.0125 0.0191 0.0256
101
0.0250 0.0379 0.0505
201
0.0500 0.0754 0.1006
401
0.1000 0.1504 0.2006
801
0.2000 0.3004 0.4005
1001
0.4000 0.6004 0.8006
6.16