EasyManua.ls Logo

HP 8753E - Page 351

HP 8753E
699 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FORWARD
I
I
I
I
E
/F
I
1
I
s
I
1
I
21A
ETF
s
I
21M
RF IN l
-
1
1
b
EDF\r
I
5
I
E
E
I
IIA
I
LF
SF
I
‘22A
I
I
l
U
T
A
I
s
I
IlM
ERF
I
s
12A
I
I
I
PORT 1
PORT 2
REVERSE
I
I
I
s
21A
I
ERR
‘22M
1
I
l
I
I
ESR
I
‘12M
ELF!
I
dEDP
.
I
1
1
l RF IN
E
TR
I
12A
I 1
I
I
I
.
E
XR
I
I
pg663d
Figure 6-46. Full Two-Port Error Model
F’igure
6-47 shows the full two-port error model equations for all four S-parameters of a
two-port device. Note that the mathematics for this comprehensive model use all forward and
reverse error terms and measured values. Thus, to perform full error-correction for any one
parameter, all four S-parameters must be measured.
Applications of these error models are provided in the calibration procedures described in
Chapter 5, “Optimizing Measurement Results.
n
6-70
Application and Operation Concepts

Table of Contents

Other manuals for HP 8753E

Related product manuals