For microstrip and other fabricated standards, the velocity factor is significant. In those cases,
the phase calculation must be divided by that factor. For example, if the dielectric constant for
a substrate is 10, and the corresponding “effective” dielectric constant for microstrip is 6.5,
then the “effective” velocity factor equals 0.39 (1
+
square root of 6.5).
Using the first equation with a velocity factor of 0.39, the initial length to test would be
1.95 cm. This length provides an insertion phase at 1000 MHz of 60 degrees; at 2000 MHz,
120 degrees (the insertion phase should be the same as the air line because the velocity factor
was accounted for when using the
iirst
equation).