TOO
MANYNESTED
SEQUENCES.
SEQ
ABORTED
Error Number You can only nest sequences to a maximum level of six. The sequence will
164
abort if you nest more than six.
TOO
MANYSEGMENTS
ORPOINTS
Error Number You can have a maximum of 30 segments or 1632 points in frequency list mode.
50
In power meter calibrations, you can have a maximum of 12 segments for
power sensor cal factors and power loss functions.
TRANSFORM,
GATENOTALLOWED
Error Number You can perform a time domain transformation
only
in linear and CW sweep
16
types.
TROUBLE!
CHECKSETUPAND
STARTOVER
Service Error Your equipment setup for the adjustment procedure in progress is not correct.
Number 115
Check the setup diagram and instructions
HP 8753E Network
Anulgm
semrice
ouide. Start the procedure again.
WAITINGFORCLEAN
SWEEP
Information
In single sweep mode, the instrument ensures that
aI.l
changes to the
Message
instrument state, if any, have been implemented before taking the sweep. The
command that you have initiated is being processed and
wiII
not be complete
until the new sweep is completed. An asterisk * is displayed in the left margin
until a complete fresh sweep has been taken.
WAITINGFORDISK
Information
This message is displayed between the start and
finish
of a read or write
Message
operation to a disk.
lo-26
Error Messages