5-26 CLARiTY Power Saving
ICE Vulcan Service Manual
Power Saving Mode = 0: None
User interface will not dim when power is applied.
Power Saving Mode = 1: Minimal
User interface will dim after 1 minute of inactivity.
Activity:
•
User touching the screen
•
USB keyboard input
•
Raising of an warning (e.g. the moment the LPA transitions from No
Error into Warning)
•
Presence of an error
Power Saving Mode = 2: High
User interface dims after 1 minute of inactivity.
User interface turns off after a further 14 minutes of inactivity.
Activity:
•
User touching the screen
•
USB keyboard input
•
Printing
•
Raising of a warning (e.g. the moment the LPA transitions from No
Error into Warning)
•
Presence of an error
Power Saving Mode = 3: Full
•
User interface dims after 1 minute of inactivity.
•
User interface turns off after a further 14 minutes of inactivity.
Activity:
•
User touching the screen
•
USB keyboard input
•
Raising of a warning (e.g. the moment the LPA transitions from No
Error into Warning)
•
Presence of an error
Note: When an error is present the screen will be at full brightness.