ICE Vulcan Service Manual
Horizontal Stand with Conveyor (U Base)
Table 6-2: Replacement Instructions (Continued)
Motor Calibration Wizard
During the LPA build, the ribbon and the printhead motors are calibrated
and the calibration
is stored in the PCB. When the PCB or the motors are
changed, the calibration data is no
longer valid and recalibration is
required.
•
When the PCB is changed, the installation wizard will start on the
LPA power ON and the
calibration process is part of this procedure.
•
When the printhead or ribbon motors are replaced, it is necessary to
carry out the
calibration procedure as explained below.