Rev AA
Maintenance 6-5
ICE Vulcan Service Manual
Calibrate Ribbon Motors
Do the following tasks to calibrate ribbon motors:
Note: Remove the label or ribbon fitted to the mandrels before starting calibration
procedure.
1
Navigate to Operations > CharacterizeRibbonMotors.
Figure 6-1: Characterize Ribbon Motors