(h) After about 6 seconds, the LOAD CAL LED goes
out and the display goes blank indicating that calibra-
tion is completed. Release the calibration button.
(i) Press LOAD BAL key. The LOAD BAL LED lights
(j) Press ENTER. After about 3 seconds, the LOAD
BAL LED goes out indicating that the balance func-
tion is completed. (Go to “Load System Balance” on
page 5-15.)
Manual Calibration of Non-Self Identifying
Load Cells
Note When manually (deadweight) calibrating a
load cell, use a precision weight that is trace-
able to a particular class as defined by the
National Bureau of Standards.
To calibrate the load weighing system manually with a
non-self identifying load cell installed:
(a) Do one of the following:
TENSION CELL
CATALOG NO.
MAXIMUM
CAPACITY
CALIBRATION
POINT
2512-136
2512-101
2512-122
50 g
500 g
5N
20 g
200 g
2N
Table 5-1. Low Capacity Load Cells Calibration Data
System
Operation
Load Calibration
5-13