EasyManua.ls Logo

Instron 4400 Series - Limits Section Functions

Instron 4400 Series
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STATUS
LEDs
A rectangular STATUS LED is located to the right of the
BREAK key and one each between the LOAD,
EXTENSION, and STRAIN MAX/MIN limit keys. When a
limit or break detection is selected to cause a crosshead
action (except OFF), the related STATUS LED lights.
When either a STOP or RETURN action occurs, the
STATUS LED will flash and stay flashing until recycled by
starting a new test.
If the current value of load, extension or strain is beyond
the assigned limit ranges, then the crosshead control key
(UP or DOWN) which would increase the out-of-range
error is disabled. If this disabled key is pressed, the
crosshead will not move and the related limit status LED
flashes. The flashing continues until the enabled key (UP
or DOWN) is pressed and the crosshead is moved in a
direction to decrease the out-of-range error.
Table 4-3. Limits Section Functions (continued)
Function of
Controls
Function Of Controls
4-15

Table of Contents