The Pretension, Excess Tension, Automatic Start, and
Automatic Release functions enable or disable their re-
spective features. They are independent, except that the
Excess Tension function will not operate unless Preten-
sion is also enabled.
The Pretension and Excess Tension levels can be set in-
dependently of their respective enable controls, but will
have no effect while disabled.
Manual Operation
The grips can be operated in a fully manual mode if Grip
Control is set to Disabled, or if all of the other functions
are disabled.
As described in the Pneumatic Grip Control manual, in
manual operation, the CLOSE foot pedal is pressed once
to close the upper grip. It is pressed again to close the
lower grip. The OPEN foot pedal opens both grips simul-
Setting Default
Grip Control Enable/Disable Disabled
Pretension Enable/Disable Disabled
Excess Tension Enable/Disable Disabled
Automatic Start Enable/Disable Disabled
Automatic Release Enable/Disable Disabled
Pretension Level 0
Excess Tension Level 0
Table 6-11. Air Kit Functions
Air Kit Option M10-94400-1
6-46